header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

IN VIVO DETERMINATION OF KNEE KINEMATICS DURING GAIT FOR SUBJECTS IMPLANTED WITH A UNICOMPARTMENTAL ARTHROPLASTY



Abstract

The objective of the present study was to analyse kinematics of subjects having a UKA during stance phase of gait, where the ACL was intact at the time of the operative procedure.

Femorotibial contact positions for nineteen subjects (15 medial UKA (MUA); 14 lateral UKA (LUA); HSS > 90, post-op > 3 yrs) were analysed using video fluoroscopy.

During stance-phase of gait, on average, subjects having a medial UKA experienced 0.8 mm of anterior motion (7.7 to – 2.3 mm), while subjects having a lateral UKA experienced −0.4 mm (0.9 to – 2.1 mm) of posterior femoral rollback (PFR). Eight of 15 subjects having a medial UKA and two out of four lateral UKA experienced PFR. Eight of 15 subjects having a medial UKA experienced normal axial rotation (average = 0.9 degrees) and one out of four subjects having a lateral UKA experienced normal axial rotation (average = −6.0 degrees).

High variability in the kinematic data for subjects experiencing an anterior slide and opposite axial rotation suggests that these subjects had an ACL that was not functioning properly and was unable to provide an anterior constraint force with the necessary magnitude to thrust the femur in the anterior direction at full extension. Progressive laxity of the ACL may occur over time, and at least in part, lead to premature polyethylene wear occasionally seen in UKA. Our results support the findings of other studies that the ACL plays a significant role in maintaining satisfactory knee kinematics, which may also, in part, contribute to UKA longevity.

The abstracts were prepared by Nico Verdoschot. Correspondence should be addressed to him at Orthopaedic Research Laboratory, Universitair Medisch Centrum, Orthopaedie / CSS1, Huispost 800, Postbus 9101, 6500 HB Nijmegen, Th. Craanenlaan 7, 6525 GH Nijmegen, The Netherlands.