HALLUX VALGUS: IS THE RADIOGRAPHIC ASSESSMENT RELIABLE?



Abstract

The five different methods of measuring hallux valgus (HVA) and intermetatarsal angles (IMA) and the diagnosis of congruency of first MTP joint were studied on 50 pre-operative standing foot radiographs, to test if these methods were reliable and the results reproducible enough to be used in a treatment algorithm for hallux valgus.

Analysis of variance (ANOVA) was used to examine the difference between the five methods and between the five observers. Kappa test was used to measure agreement in diagnosing congruency between two occasions.

The mean IMA and HVA varied significantly (p< 0.00001). The ANOVA model showed that method and observer variations were both significant for IMA; there was no significant difference between measurement methods for HVA. Congruency had good (k=0.608) intraobserver and fair (k=0.261) interobserver reliability. A second IMA measurement will lie between 4.2° less and 4.6° more than the first IMA measurement 95% of the time. A second HVA measurement will lie between 6° less and 5.6° more than the first HVA measurement 95% of the time.

Overall, there was no advantage to any of the measurement methods, although some observers were better than others. All methods had considerable inter- and intra-observer variability that makes these measurements unreliable.

The abstracts were prepared by Mr J. L. Barrie. Correspondence should be addressed to Mr J. L. Barrie, BOFSS Editor, Department of Orthopaedic Surgery, Blackburn Royal Infirmary, Blackburn, Lancashire BB2 3LR.