THE SERUM LEVELS OF COBALT AND CHROMIUM IN A DYNAMIC TOTAL HIP SYSTEM



Abstract

Introduction: The aim of this study was to compare the serum cobalt and chromium levels in patients with Oxford Universal hips and Cenator hips. Both systems are composed of cobalt chromium alloy and have modular cobalt chromium heads, however only the Oxford hip has a sliding mechanism.

Material: The serum levels of chromium and cobalt were measured in 20 patients with Oxford Universal Hip replacement (Group 1). This was compared to the serum levels of chromium and cobalt in 17 patients with Cenator Hip replacement (Group 2) and 20 patients with no implants (Group 3). Patients in Group 1 and 2 were assessed clinically and with plain radiographs. Patients were excluded if there was evidence of loosening, if they had other joint implants or any other known cobalt or chromium implant and if they had impaired renal function. All hip replacements were performed as primary procedures. Specimens were analysed by electrothermal atomization atomic absorption spectrometry.

Results: There was no statistically significant difference in serum cobalt and chromium levels between groups 1 and 2. Patients in groups 1 and 2 had statistically significant higher serum levels of chromium and cobalt than group 3 patients.

Discussion: This study demonstrates that serum levels of chromium and cobalt are elevated in patients with stable hip implants composed of this alloy but not in a normal population. Although it is not known what the long-term effects of chronic low-grade exposure to these ions are, these levels are many order of magnitude below the toxic range. Furthermore, our results show that the Oxford Universal Hip releases no more metal ions into the circulation than an implant without a sliding mechanism.

Correspondence should be addressed to 8 Martiou Str. Panorama, Thessaloniki PC:55236, Greece.