SURVIVORSHIP OF THE CHARNLEY ELITE PLUS FEMORAL STEM



Abstract

We report the results of the Charnley Elite Plus femoral stem (Ortron 90; Depuy, Leeds, United Kingdom) in multiple surgeon’s hands at a minimum of three years post implantation.

The long term results of the Charnley femoral stem have been widely documented . There have been numerous changes to the design and instrumentation of this original stem since its introduction in 1962, and the Charnley Elite Plus represents the fifth generation of this highly successful implant.

Between March 1994 and March 1998, 244 patients underwent 268 primary hip arthroplasty procedures using this particular stem. Patients were reviewed at a mean of 4.5 years (3.0 – 6.8 years) following their arthroplasty using the Oxford Hip Score and plain radiographs.

There were five revision procedures for aseptic loosening (5/268; 1.9%). Radiological assessment revealed gross radiological failure in a further 12 femoral stems (12/208; 5.8%). There was evidence of focal osteolysis with an apparently stable implant in 36 hips (17.3%).

In the best case scenario, using revision for aseptic loosening as the endpoint, the survivorship for this period is 98.1%. If radiographic failures are incorporated into this endpoint, survivorship is 93.1%. Of potential concern however, is the number of adverse features noted on the radiographs, with only 76.9% being categorised as ‘normal.’

The Charnley Elite Plus stem has undergone some fundamental design changes from the original Charnley stem and therefore clinical success should not be automatically assumed. In such circumstances we recommend regular clinical and radiographic follow-up of patients who have have undergone total hip arthroplasty with this particular femoral stem.

The abstracts were prepared by Mr Richard Buxton. Correspondence should be addressed to him at Bankton Cottage, 21 Bankton Park, Kingskettle, Cupar, Fife KY15 7PY, United Kingdom