IATROGENIC CAUDA EQUINA SYNDROME SECONDARY TO IMPLANTABLE HAEMOSTATS



Abstract

Meticulous haemostasis not only improves the operative field facilitating spinal surgery, but also diminishes chances of post-operative neurological complications from a compressive haematoma. Since being introduced in the 1940’s, implantable haemostats have proven a useful adjunct in achieving haemostasis with relatively few complications. However, their use in spaces bounded by bony architecture can lead to compressive effects on neurological structures.

We present three cases of post-operative cauda equina syndrome – two cases following surgery for lumbar disc herniation and one case following surgery for lumbar canal stenosis. In each case, implantable haemostats were utilised to control haemorrhage for complications during the surgery. All three patients underwent urgent exploration, which revealed cauda equina compression from clot organised around the haemostat. Neurological recovery was variable.

We recommend careful attention to intra-operative haemostasis. Although haemostats can assist in achieving haemostasis, we caution against leaving them in situ.

The abstracts were prepared by Mr Richard Buxton. Correspondence should be addressed to him at Bankton Cottage, 21 Bankton Park, Kingskettle, Cupar, Fife KY15 7PY, United Kingdom