MODE OF WEAR OF TIBIAL POLYETHYLENE SURFACE IN RHEUMATOID CASES WITH YOKOHAMA MEDICAL CERAMIC TKA



Abstract

Wear of UHMWPE and Clinical results of bioceramic total knee replacement have not been well reported yet. The ultra high molecular weight polyethylene surfaces of the tibial components were examined in 3 retrieved knees with non-infectious loosening, and were almost normal in appearance with only minor scratch lines 33 to 59 months after the initial arthroplasty. Bioceramic total knee arthroplasty was concluded to show satisfactory results except for the initial several cases done with rather poor surgical techniques. The wear of UHMWPE surface in ceramic knee was observed to be much less and milder than that of metal prosthesis.

Total knee arthroplasty (cementless) using Yokohama Medical Ceramic Knee was performed in 64 knees and, excluding 4 knees with the prostheses retrieved, was evaluated in 60 knees of 47 patients.

There ware 1 man (1 knee) and 31 women (44 knees) who had rheumatoid arthritis with a mean age of 56.8±11.9 years, and 2 men (2 knees) and 11 women (13 knees) with osteoarthritis with a mean age of 70.6±6.9 years. The mean follow-up duration was 48.1±9.2 months ranging from 33 to 60 months.

The maximum knee flexion was 115±24 degrees before and 104±20 degrees after arthroplasty in the rheumatoid group, and 107±40 degrees before and 101±26 degrees after arthroplasty in the osteoarthritis group.

Clinical evaluation using The Hospital for Special Surgery Knee Criteria showed 7 knees as Excellent, 34 as Good, 7 as Fair and 12 as Poor after surgery. Complications consisted of infection (1 case), tibial plateau fracture (1), avulsion fracture of the tibial tuberosity (1) and patellar dislocation (1)