header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

EXTENSIVELY POROUS COATED STEMS: THE CAT’S WHISKERS!



Abstract

Long-term success of cementless femoral revision is dependent on achieving initial axial and torsional stability by maximising canal fill at the time of implantation. We report on a minimum 10 years clinical and radiographic follow-up of 170 patients with extensively coated cementless revision femoral components.

With a range of follow-up of 10 to 16 years and a mean of 13.2 years, a survivorship of greater than 95% was reported. Clinically, the average Postel-D’Aubigne pain and walking score improved from a preoperative score of 5.4 points to 10.8 points postoperatively. Eightytwo percent of the hips had radiographic evidence of a bone-ingrown prosthesis and 13.9% had evidence of stable fibrous fixation. Four percent of stems were unstable as seen on radiographs. Six stems were revised to larger extensively coated stems and one stem is causing pain and is unstable but has yet to be revised. The overall mechanical failure rate was 4.1%. Stress shielding was greatest in patients with stems larger than 16.5 mm and in osteoporotic bone (Dorr Type C). Nine percent of patients had significant thigh pain including all of the patients with unstable stems.

In the presence of bone loss in the proximal metaphyseal region of the femur, fixation of the femoral component is predictable when optimising prosthetic-bone fit in the diaphyseal region of the femur using an extensively porous coated femoral component.

The abstracts were prepared by Mrs Dorothy L. Granchi, Course Coordinator. Correspondence should be addressed to her at PMB 295, 8000 Plaza Boulevard, Mentor, Ohio 44060, USA.