Results of uncemented porous-coated anatomic total hip replacement



Download PDF

Abstract

We reviewed a consecutive series of 241 uncemented, porous-coated anatomic (PCA) hip replacements at an average follow-up of five years (2 to 9). Of these, 32 had failed (13%), 26 at the acetabular component (11%) and six at the femoral component (2%). Acetabular failure was associated with local osteolysis and excessive polyethylene wear in 20 cases: in these histological examination showed giant macrophages incorporating numerous particles of high-density polyethylene. The femoral failures were related to a poor intramedullary fit with subsequent subsidence. Using the recommendation for revision as the end point, the cumulative survival rate for prostheses was 91% at six years (95% CI +/- 6%), 73% (+/- 11%) at seven years, and 57% (+/- 20%) at eight years. The result of uncemented PCA hip replacement is satisfactory up to six years, but then increasing failure of the acetabular component appears to be due to polyethylene wear, leading to osteolysis, loosening and component migration. At first, failure is often asymptomatic; routine follow-up of uncemented hip replacement is essential, especially after five years.