header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Failure of acetabular cups influenced by marker wire anchorage



Download PDF

Abstract

Failure of an acetabular cup is uncommon and has been attributed to wear or creep, trauma or bony irregularities in the acetabulum. We report ten cases in which fracture of the cup occurred at the site of drill holes used to anchor the marker wire. The role of such indentations as stress raisers has not been previously reported; we suggest that deep indentations or grooves should not be placed in the most highly stressed areas and that the cup thickness should allow for predicted wear rates.

For access options please click here