Wear of high-density polyethylene on bone and cartilage



Download PDF

Abstract

Wear of high-density polyethylene on bone and cartilage has resulted in a large volume of plastic particles being shed into the two knees and two hips studied. The giant-cell foreign-body reaction of the synovium may not be sufficient to cope with the amount of debris presented and the destruction of the endosteal bone in one hip, caused by the wear particles and movement of the prosthesis, has made revision impossible. Articulation of high-density polyethylene against bone or cartilage either by design or by the failure of alignment of the component must be avoided.