Hip

Fixation of the short global tissue-sparing hip stem

Two-year follow-up results of a randomized controlled clinical and Roentgen stereophotogrammetric analysis study



Download PDF

Abstract

Aims

Short, bone-conserving femoral components are increasingly used in total hip arthroplasty (THA). They are expected to allow tissue-conserving implantation and to render future revision surgery more straightforward but the long-term data on such components is limited. One such component is the global tissue-sparing (GTS) stem. Following the model for stepwise introduction of new orthopaedic implants, we evaluated early implant fixation and clinical outcome of this novel short-stem THA and compared it to that of a component with established good long-term clinical outcome.

Methods

In total, 50 consecutive patients ≤ 70 years old with end-stage symptomatic osteo-arthritis were randomized to receive THA with the GTS stem or the conventional Taperloc stem using the anterior supine intermuscular approach by two experienced hip surgeons in two hospitals in the Netherlands. Primary outcome was implant migration. Patients were followed using routine clinical examination, patient reported outcome using Harris Hip Score (HHS), Hip Disability And Osteoarthritis Outcome Score (HOOS), EuroQol five-dimension questionnaire (EQ5D), and Roentgen Stereophotogrammetric Analysis (RSA) at three, six, 12, and 24 months. This study evaluated the two-year follow-up results.

Results

In addition to the initial migration pattern of distal migration (subsidence, Y-translation) and retroversion (Y-rotation) also exhibited by the Taperloc stem, the GTS stem showed an initial migration pattern of varization (X-translation combined with Z-rotation) and posterior translation (Z-translation). However, all components stabilized aside from one Taperloc stem which became loose secondary to malposition and was later revised. Clinical outcomes and complications were not statistically significantly different with the numbers available.

Conclusion

A substantially different and more extensive initial migration pattern was seen for the GTS stem compared to the Taperloc stem. Although implant stabilization was achieved, excellent long-term survival similar to that of the Taperloc stem should not be inferred. Especially in the absence of clinically proven relevant improvement, widespread usage should be postponed until long-term safety has been established.

Cite this article: Bone Joint J 2020;102-B(6):699–708.


Correspondence should be sent to Marc J. Nieuwenhuijse; E-mail: